Dimensions / Weight
W | H | D | |
---|---|---|---|
Dimensions | 430 mm (16.93 inch) | 177 mm (6.97 inch) | 370 mm (14.57 inch) |
AC/DC Hipot Tester
TOS5101
TOS5101
470,000円 (税込:517,000円)
- Compliant with various safety standards
- Equipped with remote control function
- DC output also available
- Various signal output functions
- Digital voltmeters and ammeters
- Automatic discharge function (during DC operation)
- Digital timer
- Provided with zero turn-on switch
- Window comparator type employed for PASS/FAIL judgement
*Discontinued and only available while supplies last
This product generates high voltage! The use of this product requires knowledge of electricity.Improper handling or operation of this product can be very dangerous, so be sure to read the instruction manual carefully and use this product properly.
Overview
Exclusive model of TOS Series having AC, DC10 kV output
TOS5101 is designed exclusively for withstand voltage testing of electronic equipment and components conforming to various safety standards. The use of a high luminance, large fluorescent display tube for the display enables data including measured values, status and judgment results to be extremely legible. The PASS/FAIL function employs a window comparator method that enables TOS5101 to make fail judgment of current leakage over the upper reference value and below the lower reference value which can be set on the front panel.
Thus, highly reliable testing can be performed including that for test lead disconnection and defective contact. In addition, in order to prevent erroneous operation and accidents, the TOS5101 is also equipped with a Key Lock function and Interlock function, a high- voltage output terminal having a narrowed insertion port, a large DANGER lamp, and an automatic discharge function (during DC operation) that removes charge from the test piece. These features give the TOS5101 a high degree of safety and reliability.
*In general, when the capacitance of DUT has a voltage dependence (such as a “High-dielectric constant ceramic capacitor”), please take a caution that the waveform distortion may occurs.
Dimensions
Specifications
Output block
Applied Voltage | 0 to 5/ 0 to 10 kV AC and DC |
---|
AC
Maximum Rated | 500 VA / 10 kV, 50 mA | |
---|---|---|
Waveforms | Commercial power supply waveform | |
Voltage Regulation | Max. 15% (for max. rated load to no load) | |
Switching | Use of a zero turn-on switch |
DC
Maximum Rated | 50W / 10 kV, 5 mA | |
---|---|---|
Ripple | 100 Vp-p typ. at 10 kV, no load 200 Vp-p typ. at max. rated output | |
Voltage Regulation | Max. 3% (for max. rated load to no load) |
Output voltmeter
Analog | Scale | 10 kV full scale , AC/DC |
---|---|---|
Class | JIS Class 2.5 | |
Accuracy | ±5% of full scale | |
AC Indication | Mean value response / rms value scale | |
Digital | Full Scale | 5 kV/ 10 kV full scale |
Accuracy | ±1.5% of full scale | |
AC Response | Mean value response / rms value display |
Ammeter
Digital | Accuracy | ±(5% + 20μA) of upper cutoff current |
---|---|---|
AC Response | Mean value response / rms value display |
Pass/fail Judgement Function
Type of Judgement | Window comparator type
| |
---|---|---|
Upper cutoff current setting range | AC: 0.1 to 55 mA DC: 0.1 to 5.5 mA | |
Lower cutoff current setting range | AC: 0.1 to 55 mA DC: 0.1 to 5.5 mA | |
Judgment accuracy | ±(5% of upper cutoff current + 20μA) | |
Current detection method | Integration of current absolute value followed by comparison with reference value. | |
Calibration | Calibration is performed with the rms value of the sine wave, using a pure resistance load. | |
No-load output voltage required for detection | Approx. 970 V when set to 50 mA AC | |
Approx. 160 V when set to 5 mA DC |
Test Time
Setting range | 0.5 to 999 sec (±10 ms) (timer-off function provided) | |
---|---|---|
Accuracy | ±20ms |
RS232C interface
Connector | – | |
---|---|---|
Protocol | – | |
Function | – |
AC line power requirements
Line Voltage | 100V±10%, 50/60 Hz (Nominal voltages of 110V, 120V, 220V, 230V and 240V available as factory options.) |
---|
Power Requirements
for line voltage of 100 V | Max. 50 VA under no-load conditions / Approx. 600 VA at rated load |
---|---|
for line voltage of 100 V to 200 V | Max. 50 VA under no-load conditions / Approx. 600 VA at rated load |
for line voltage of 220 V to 240 V | Max. 50 VA under no-load conditions / Approx. 610 VA at rated load |
Electromagnetic compatibility (EMC) *1
Conforms to the requirements of the following standard.
EMC Directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
Under following conditions 1. Used HV test leadwires which is supplied.
2. No discharge in testing.
3. Used the shielded cable which length is less than three meters when the SIGNAL I/O is used.
*1 Not applicable to custom order models.
Environment
Specification range | 5 °C to 35°C(41 to 95˚F) / 20 %rh to 80 %rh |
---|---|
Operable range | 0 °C to 40°C(32 to 104˚F) / 20 %rh to 80 %rh |
Storage range | -20 °C to 70 °C (–4 to 158˚F) / 80 %rh or less |
Dimensions (maximum)
430[16.9 inch] W × 177[6.97 inch] (195[7.68 inch]) H × 370[14.6 inch] (450[17.7 inch]) D mm
Weight
for line voltage of 100 V | Approx. 21 kg (Approx. 46.30 lbs) |
---|---|
for line voltage of 100 V to 200 V | Approx. 23 kg (Approx. 50.70 lbs) |
for line voltage of 220 V to 240 V | Approx. 24 kg (Approx. 52.91 lbs) |
Accessories
High-voltage test lead | TL01-TOS (max.allowablevoltage: 5 kV /1.5m) TL03-TOS (max.allowablevoltage: 10 kV /1.5m) |
---|---|
Dimensions | 14-pin amphenol plug (assembled) |
Dimensions(mm)
Pin Configuration for the SIGNAL I/O Connector
Options
High-voltage test leads for withstand voltage test
Note: Use with TOS5101 is possible within the cable rating.