AC/DC Hipot Tester

TOS5101

TOS5101

470,000 (税込:517,000

  • Compliant with various safety standards
  • Equipped with remote control function
  • DC output also available
  • Various signal output functions
  • Digital voltmeters and ammeters
  • Automatic discharge function (during DC operation)
  • Digital timer
  • Provided with zero turn-on switch
  • Window comparator type employed for PASS/FAIL judgement

*Discontinued and only available while supplies last

This product generates high voltage! The use of this product requires knowledge of electricity.Improper handling or operation of this product can be very dangerous, so be sure to read the instruction manual carefully and use this product properly.


カート機能が動くまでの

Overview

Exclusive model of TOS Series having AC, DC10 kV output

TOS5101 is designed exclusively for withstand voltage testing of electronic equipment and components conforming to various safety standards. The use of a high luminance, large fluorescent display tube for the display enables data including measured values, status and judgment results to be extremely legible. The PASS/FAIL function employs a window comparator method that enables TOS5101 to make fail judgment of current leakage over the upper reference value and below the lower reference value which can be set on the front panel.

Thus, highly reliable testing can be performed including that for test lead disconnection and defective contact. In addition, in order to prevent erroneous operation and accidents, the TOS5101 is also equipped with a Key Lock function and Interlock function, a high- voltage output terminal having a narrowed insertion port, a large DANGER lamp, and an automatic discharge function (during DC operation) that removes charge from the test piece. These features give the TOS5101 a high degree of safety and reliability.

*In general, when the capacitance of DUT has a voltage dependence (such as a “High-dielectric constant ceramic capacitor”), please take a caution that the waveform distortion may occurs.

Dimensions

Specifications
Open in a new windowZOOM

Output block

Applied Voltage0 to 5/ 0 to 10 kV AC and DC

AC

Maximum Rated500 VA / 10 kV, 50 mA
WaveformsCommercial power supply waveform
Voltage RegulationMax. 15% (for max. rated load to no load)
SwitchingUse of a zero turn-on switch

DC

Maximum Rated50W / 10 kV, 5 mA
Ripple100 Vp-p typ. at 10 kV, no load 200 Vp-p typ. at max. rated output
Voltage RegulationMax. 3% (for max. rated load to no load)

Output voltmeter

AnalogScale10 kV full scale , AC/DC
ClassJIS Class 2.5
Accuracy±5% of full scale
AC IndicationMean value response / rms value scale
DigitalFull Scale5 kV/ 10 kV full scale
Accuracy±1.5% of full scale
AC ResponseMean value response / rms value display

Ammeter

DigitalAccuracy±(5% + 20μA) of upper cutoff current
AC ResponseMean value response / rms value display

Pass/fail Judgement Function

Type of JudgementWindow comparator type
  • ●FAIL judgement
  • *When current detected above upper cutoff current
  • *When current detected below lower cutoff current (FAIL signal generated when FAIL judgement made)
  • ● PASS judgement *When set time has elapsed and no abnormality is detected
Upper cutoff current setting rangeAC: 0.1 to 55 mA DC: 0.1 to 5.5 mA
Lower cutoff current setting rangeAC: 0.1 to 55 mA DC: 0.1 to 5.5 mA
Judgment accuracy±(5% of upper cutoff current + 20μA)
Current detection methodIntegration of current absolute value followed by comparison with reference value.
CalibrationCalibration is performed with the rms value of the sine wave, using a pure resistance load.
No-load output voltage required for detectionApprox. 970 V when set to 50 mA AC
Approx. 160 V when set to 5 mA DC

Test Time

Setting range0.5 to 999 sec (±10 ms) (timer-off function provided)
Accuracy±20ms

RS232C interface

Connector
Protocol
Function

AC line power requirements

Line Voltage100V±10%, 50/60 Hz (Nominal voltages of 110V, 120V, 220V, 230V and 240V available as factory options.)

Power Requirements

for line voltage of 100 VMax. 50 VA under no-load conditions / Approx. 600 VA at rated load
for line voltage of 100 V to 200 VMax. 50 VA under no-load conditions / Approx. 600 VA at rated load
for line voltage of 220 V to 240 VMax. 50 VA under no-load conditions / Approx. 610 VA at rated load

Electromagnetic compatibility (EMC) *1

Conforms to the requirements of the following standard.
EMC Directive 2004/108/EC

EN 61326-1
EN 61000-3-2
EN 61000-3-3
Under following conditions 1. Used HV test leadwires which is supplied.

2. No discharge in testing.
3. Used the shielded cable which length is less than three meters when the SIGNAL I/O is used.

*1 Not applicable to custom order models.

Environment

Specification range5 °C to 35°C(41 to 95˚F) / 20 %rh to 80 %rh
Operable range0 °C to 40°C(32 to 104˚F) / 20 %rh to 80 %rh
Storage range-20 °C to 70 °C (–4 to 158˚F) / 80 %rh or less

Dimensions (maximum)

430[16.9 inch] W × 177[6.97 inch] (195[7.68 inch]) H × 370[14.6 inch] (450[17.7 inch]) D mm

Weight

for line voltage of 100 VApprox. 21 kg (Approx. 46.30 lbs)
for line voltage of 100 V to 200 VApprox. 23 kg (Approx. 50.70 lbs)
for line voltage of 220 V to 240 VApprox. 24 kg (Approx. 52.91 lbs)

Accessories

High-voltage test leadTL01-TOS (max.allowablevoltage: 5 kV /1.5m) TL03-TOS (max.allowablevoltage: 10 kV /1.5m)
Dimensions14-pin amphenol plug (assembled)

Dimensions(mm)


TOS5101 dimensions

Pin Configuration for the SIGNAL I/O Connector


Pin Configuration for the SIGNAL I/O Connector

Product Catalog