Constant voltage temperature coefficient | 100 p.p.m./°C (standard value) |
---|---|
Constant current temperature coefficient | 300 p.p.m./°C (standard value) |
Transient response time | 50 μs: Time required for the output voltage to return to a value less than 0.05 % of the rated value + 10 mV, against a fluctuation of 5 % to 100 % of the output current. |
Ripple noise | Using an AC voltmeter having a range of 5 Hz to 1 MHz, ±3 dB, indicated in mean value and effective value measurement is performed with either a positive or negative output terminal connected to the ground. |
Indication meters | Voltmeter indication error : ±(0.5 % rdg + 2 digits) at 23 °C ±5 °C |
Voltmeter max. indication digits : 199.9 (Note: 19.99 for the PAN16-10A/PAN16- 18A/PAN16-30A/PAN16-50A models, and 1999 for the PAN250-2.5A/PAN250-4.5A/ PAN600-2A) | |
Amperemeter indication error : ±(1 % rdg + 5 digits) at 23 °C ±5 °C | |
Amperemeter max. indication digits : 19.99 (Note: 1.999 for the PAN110-1.5A/ PAN160-1A, and 199.9 for the PAN16-30A/ PAN16-50A/PAN35-20A/PAN35-30A/PAN60- 20A) | |
Grounding | Plus or minus terminal can be grounded |
Isolation voltage | ±250V DC (However, ±500 VDC for the PAN110-1.5A/PAN110-3A/PAN110-5A/ PAN110-10A/PAN160-1A/PAN160-2A/ PAN160-3.5A/PAN250-2.5A/PAN250-4.5A.) (However, PAN350-3.5A and PAN600-2A are ±1000V DC) |
Insulation resistance | Across input side and chassis: Greater than 30 MΩ at 500 VDC |
Across output side and chassis: Greater than 20 MΩ at 500 VDC (For PAN350-3.5A and PAN600-2A, this is 1000 V DC, 20 MΩ, or higher.) | |
Withstand voltage | Nothing abnormal should occur at 1500 VAC, 1 min. |
Operating temperature range | 0° to 40°C |
Operating humidity range | 10 to 90 % RH |
Cooling system | Forced air cooling using fan |
Constant voltage operation | Green LED |
Constant current operation | Red LED |
Protection devices | Constant voltage, constant current automatic crossover |
Overvoltage protector (OVP) (10 % to 110 % of rated output voltage) | |
Overcurrent protection circuit (Approx. 110 % of rated output current) | |
Overvoltage protection circuit (Smoothing electrolytic capacitor for the rectification circuit) | |
Overheat protection circuit (OHP) (Semiconductor cooling heat-sink [100 °C]) | |
Thermal fuse (Main- or sub-transformer) | |
Input/output fuse | |
Input surge absorber | |
About input voltage | If you require a power supply voltage other than 100V, please contact us at the time of order placement. (Factory option) |
Dimensions (maximum dimension) | Type 0: 106(4.17˝)W ✕ 140(5.51˝)H ✕ 400(15.75˝)Dmm(inch) Type I 2: 210(8.27˝)W ✕ 140(5.51˝)H ✕ 350(13.78˝)Dmm(inch) Type I 3: 210(8.27˝)W ✕ 140(5.51˝)H ✕ 400(15.75˝)Dmm(inch) Type II: 430(16.93˝)W ✕ 160(6.3˝)H ✕ 400(15.75˝)Dmm(inch) |