Alarm 1 (high urgency)
Name | Operation |
---|---|
Overvoltage detection (OVP) | Load off |
Reverse-connection detection (Reverse) | |
Overheat detection, overcurrent detection of the front-panel DC INPUT terminals (OTP/Front) | |
Alarm input detection (External) | |
Parallel operation anomaly detection |
Alarm 2 (low urgency)
Name | Mode | Operation |
---|---|---|
Overcurrent protection (OCP) | CR, CV, CP | Load off or limit |
Overpower protection (OPP) | CC, CR, CV, ARB | |
Undervoltage protection (UVP) | CC, CR, CP, ARB | Load off, limit, or activation off |
Watchdog protection (WDP) | All | Load off |
This function limits the current (UVPL) or turns off the load (UVPT) when the voltage applied to the product becomes equal to or less than the UVP setting. You can set this regardless of whether the load is on or off.
Trip | Turns the load off. The setting display changes to UVPT. |
---|---|
Limit | Limits the voltage so as not to become equal to or less than the set value. The setting display changes to UVPL. |