Item/Model | 400W | 800W | 1200W | 2000W | |
---|---|---|---|---|---|
Overvoltage protection (OVP) | Turns the output off *1, displays OVP, and lights ALM | ||||
Selectable range | 10 % to 112 % of the rated output voltage | ||||
Setting accuracy | ±(1.5 % of rating) | ||||
Overcurrent protection (OCP) *2 | Turns the output off *1, displays OCP, and lights ALM | ||||
Selectable range | 10 % to 112 % of the rated output voltage | ||||
Setting accuracy | ±(3 % of rating) | ||||
Front-panel output terminal overcurrent protection (FOCP)*3 | Turns the output off *1, displays FOCP, and lights ALM | ||||
Value (fixed) | 11 A (TYP) | ||||
Undervoltage limit (UVL) | Cannot be set to a value less than or equal to the set voltage | ||||
Selectable range | 0 % to 105 % of the rated output voltage | ||||
Overheat protection (OHP) | Turns the output off, displays OHP, and lights ALM | ||||
Incorrect sensing connection protection (SENSE) | Turns the output off, displays SENS, and lights ALM | ||||
Low AC input protection (AC-FAIL) | Turns the output off *4, displays AC, and lights ALM | ||||
Shutdown (SD) | Turns the output off *1, displays SD, and lights ALM | ||||
Power limit (POWER LIMIT) | ALM blinking | ||||
Value (fixed) | Approx. 105% of the rated output power | ||||
Communication monitoring (watchdog) | Turns the output off, displays WDOG, and lights ALM | ||||
Master-slave parallel operation protection (PRL ALM) | Turns the output off *1, displays PRL, and lights ALM |
*1. Output off or breaker trip on the 2000 W model. *2. This does not protect against the discharge current peak that is generated from the capacitors inside the PWR-01 output section when the load is changed suddenly. *3. Available on models with a maximum settable current of 11 A or more. If the OCP value is less than the FOCP value, the OCP value takes precedence. *4. Auto recovery after eliminating the cause of the alarm is selectable.