TOS9300_Leakage Current Test-Judgment function

ItemTOS9303LC
Behavior based on judgmentJudgment starts after the judgment delay (Judge Delay). Buzzer volume level can be set in the range of 0 (OFF) to 10 for pass and fail separately. In an auto test, the buzzer is valid only for the judgment that takes place at the end of the program.
UPPER FAILJudgment methodUPPER FAIL results when a current greater than or equal to the upper limit (Upper) is detected.
Display“U-FAIL” is displayed.
BuzzerOn
SIGNAL I/OThe Upper-FAIL signal is generated continuously until a STOP signal is received.
Lower FAILJudgment methodLOWER FAIL results when a current less than or equal to the lower limit (Lower) is detected.
Display“L-FAIL” is displayed.
BuzzerOn
SIGNAL I/OThe Lower-FAIL signal is generated continuously until a STOP signal is received.
PASSJudgment methodPASS judgment is made if U-FAIL or L-FAIL has not occurred when the test time elapses.
Display“PASS” is displayed.
BuzzerOn (fixed to 50 ms)
SIGNAL I/OThe PASS signal is generated for the length of time specified by the Pass Hold setting. If Pass Hold is set to Infinity, the PASS signal is generated continuously until a STOP signal is received.
Upper Setting rangeRANGE1DC, RMS: 0.1 μA(min.) to 200 μA(max), Peak: 0.1 μA(min.) to 282 μA(max)
RANGE2DC, RMS: 15.1 μA(min.) to 2.00 mA(max), Peak: 21.3 μA(min.) to 2.83 mA(max)
RANGE3DC, RMS: 151 μA(min.) to 20.00 mA(max), Peak: 213 μA(min.) to 28.3 mA(max)
RANGE4DC, RMS: 1.51 mA(min.) to 100 mA(max), Peak: 2.13 mA(min.) to 100 mA(max)
Lower Setting rangeA value that is -1 digit from the upper setting range.
Judgment accuracyConforms to total accuracy(Read “reading” as “upper setting” of total accuracy.)

Related specifications : TOS9300 Series