Digit Multimeter

DME1600

DME1600

134,000 (税込:147,400

  • Resolution : 6 1/2 digit
  • Display : 5 x 7 dot matrix VFD, dual display with 3-color
  • Basic measurement function
    • DC voltage : 0.1 V, 1 V, 10 V, 100 V, 1000 V
    • AC voltage : 0.1 V, 1 V, 10 V, 100 V, 750 V
    • DC current : 10 mA, 100 mA, 1 A, 3 A
    • AC current : 1 A, 3 A
    • 2-wire / 4-wire resistance : 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ
    • Frequency : 3 Hz to 300 kHz
    • Continuity test
    • Diode test
    • Temperature test
  • Built-in USB Interface (GPIB Interface: selected model)
    *Model with GPIB Interface : DME1600GC

LAN LXIUKCA
Login to myKIKUSUI / What is myKIKUSUI?
カート機能が動くまでの

Overview


Essential device of “Electronic Measurement”
Supporting basic measurement with variety of options

The DME1600 is a digital multi-meter with a resolution of 6 1/2 digit. It can be performed up to 2000 times per second at the setting condition of 4 1/2 digit as fastest measurement, and it can measures 50 times per second when it is set for the 6 1/2 digit. The DME1600 offers fully function of measurement for the voltage, current, resistance, frequency and temperature which can be used various application of measurement and evaluation in design, development and debugging of electronics devices. The DME1600 provides USB and GPIB interface as standard feature for automated measurement besides manual operation. Furthermore, the DME1600 offers wide range of options such as 20-channel multi-point scanner card supporting the basic measurement.

Specifications

DC Characteristics
ZOOM

Accuracy

  • ±(% of reading + % of range)
  • The specifications are for the following conditions: 6 1/2 digit resolution, minimum two-hour warm up, and auto trigger mode.
  • Use the null function for the 2-wire / 4-wire resistance measurement method.

DC voltage

RangeResolutionInput resistance1 year (23 °C ± 5 °C)
100.0000 mV0.1 μV> 10 GΩ0.0050+0.0035
1.000000 V1.0 μV> 10 GΩ0.0040+0.0007
10.00000 V10 μV> 10 GΩ0.0035+0.0005
100.0000 V100 μV10 MΩ0.0045+0.0006
1000.000 V1 mV10 MΩ0.0045+0.0010

DC current

RangeResolutionInput resistance1 year (23 °C ± 5 °C)
10.00000 mA10 nA5.1 Ω0.050+0.020
100.0000 mA100 nA5.1 Ω0.050+0.005
1.000000 A1 μA0.1 Ω0.100+0.010
3.000000 A10 μA0.1 Ω0.120+0.020

Resistance

RangeResolutionInput resistance1 year (23 °C ± 5 °C)
100.0000 Ω100 μΩ1 mA0.010+0.004
1.000000 kΩ1 mΩ1 mA0.010+0.001
10.00000 kΩ10 mΩ100 μA0.010+0.001
100.0000 kΩ100 mΩ10 μA0.010+0.001
1.000000 MΩ1 Ω5 μA0.010+0.001
10.00000 MΩ10 Ω500 nA0.040+0.001
100.0000 MΩ100 Ω500 nA||10 MΩ0.800+0.010

Diode test

RangeResolutionTest current1 year (23 °C ± 5 °C)
1.0000 V10 μV1 mA0.010+0.020

Continuity test

RangeResolutionTest current1 year (23 °C ± 5 °C)
1 kΩ10 mΩ1 mA0.010+0.030

Measurement characteristics

ItemSpecifications
DC voltage measurement : OverrangePermits voltages that are up to 20 % over the range except when the 1000 V range is in use
DC voltage measurement : Input bias currentLess than 30 pA (at 25 °C)
DC voltage measurement : Input voltage protection1000 V for all ranges
DC current measurement : OverrangePermits currents that are up to 20 % over the range except when the 3 A range is in use
Resistance measurement : Maximum test lead resistance10 Ω (100 Ω range)
100 Ω (1 kΩ range)
1 kΩ (other ranges)
Resistance measurement : Input voltage protection1000 V for all ranges
AC Characteristics
ZOOM

Accuracy

  • ±(% of reading + % of range)
  • The specifications are for the following conditions: 6 1/2 digit resolution, minimum two-hour warm up, and slow AC filter (3 Hz to 300 kHz bandwidth).
  • Measured using a sine wave input whose amplitude is greater than 5% of range.
  • For the input range from 1 % to 5 %, add 0.1 % of the range (when it is less than 50kHz) or adding 0.13 % of the range (when it is from 50 kHz to 100 kHz)

AC voltage (true rms value)

RangeResolutionFrequency1 year (23 °C ± 5 °C)
100.0000 mV0.1 μV3 Hz to 5 Hz1.00+0.04
5 Hz to 10 Hz0.35+0.04
10 Hz to 20 kHz0.06+0.04
20 kHz to 50 kHz0.12+0.05
50 kHz to 100 kHz0.60+0.08
100 kHz to 300 kHz4.00+0.50
1.000000 V to 750.000 V1.0 μV to 1 mV3 Hz to 5 Hz1.00+0.03
5 Hz to 10 Hz0.35+0.03
10 Hz to 20 kHz0.06+0.03
20 kHz to 50 kHz0.12+0.05
50 kHz to 100 kHz0.60+0.08
100 kHz to 300 kHz4.00+0.50

AC current (true rms value)

RangeResolutionFrequency1 year (23 °C ± 5 °C)
1.000000 A1 μA3 Hz to 5 Hz1.00+0.04
5 Hz to 10 Hz0.30+0.04
10 Hz to 5 kHz0.10+0.04
3.000000 A10 μA3 Hz to 5 Hz1.10+0.06
5 Hz to 10 Hz0.35+0.06
10 Hz to 5 kHz0.15+0.06

Measurement characteristics

ItemSpecifications
OverrangeCapable 20 % of the over rage excluding “750 V RMS range”
Measurement frequency750 V RMS Range is limited to 100 kHz
Frequency and period characteristics
ZOOM

Accuracy

  • ±(% of reading)
  • The specifications are for the following conditions: 6 1/2 digit resolution and minimumtwo-hour warm up.
RangeFrequency1 year (23 °C ± 5 °C)
100 mVrms to 750 Vrms3 Hz to 5 Hz0.10
5 Hz to 10 Hz0.05
10 Hz to 40 Hz0.03
40 Hz to 300 kHz0.01

Measurement characteristics

ItemSpecifications
OverrangePermits voltages that are up to 20 % over the range except when the 750 Vrms range is in use
Measurement frequency750 V RMS Range is limited to 100 kHz
General specifications
ZOOM
ItemSpecifications
Input voltage range100 Vac/120 Vac/220 Vac/240 Vac ± 10 % (single phase)
Input frequency range50 Hz/60 Hz±10 %
Power25 VAmax
Operating temperature range0 °C to 50 °C
Operating humidity rangeUp to 80 % rh (0 °C to 31 °C, non condensing)
Storage temperature range-40 °C to 70 °C
Operating altitudeUp to 2000 m
Dimensions/Weight224(8.82)W×113(4.45)H×373(14.69)D mm(inch) Approx. 3.7 kg (8.2 lb)
InterfaceUSB 2.0, GPIB (factory option)
Safety *1Complies with the requirements of the following directive and standard. Low Voltage Directive 2014/35/EU EN 61010-1 (Class I, Pollution degree 2), EN 61010-2-030
Electromagnetic compatibility (EMC) *1 *2Complies with the requirements of the following directive and standard. EMC Directive 2014/30/EU EN 61326-1 (Class B ) EN 55011 (Class B, Group 1) EN 61000-3-2, EN 61000-3-3
Accessories“Power cord” 1 pc. (with three-pronged plug), “Standard test leads” (1 red, 1 black), “USB cable” 1pc., “Fuse”(spare) 1pc., “CD-ROM”*1pc., “Packing list,safety precautions” (1 English, 1 Japanese) *Contains the user’s manual and the remote interface manual.

*1 Does not apply to specially made or modified DME1600s. *2 Limited to products that have the CE/UKCA mark on their panels. *3 This is a Class I equipment. Be sure to ground this product’s protective conductor terminal. The safety of this product is only guaranteed when the product is properly grounded. *4 Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasional temporary conductivity caused by condensation. *5 This is a Class A equipment. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *6 This is a Group 1 equipment. This product does not generate and/or use intentionally radio-frequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *7 Contains the User’s Manual and the Remote Interface Manual.

Product Catalog

Icon

    Operation Manuals

    DME1600
    Icon

      Please feel free to contact us with any questions or concerns.

      Download a Product Catalog View List
      Request a Quotation Contact Us
      close

      Search

      Entering the model number, capacity, etc.