TOS9300_Leakage Current Test-Measurement section

ItemTOS9303LC
Measurement range *1Range1DC, RMS: 1.00 μA(min.) to 200.00 μA(max), Peak: 1.00 μA(min.) to 282.00 μA(max)
Range2DC, RMS: 12.50 μA(min.) to 2000.0 μA(max), Peak: 17.50 μA(min.) to 2830.0 μA(max)
Range3DC, RMS: 125.0 μA(min.) to 20.000 mA(max), Peak: 175.0 μA(min.) to 28.300 mA(max)
Range4DC, RMS: 1.250 mA(min.) to 100.00 mA(max), Peak: 1.750 mA(min.) to 100.00 mA(max)
Range switchingAuto or Fix selectable. If a measurement falls outside the measurement range of each range, the measured value blinks as a warning.
AutoThe range is set automatically according to the measurements.
FixFor TC and PCC measurements, the measurement range is selected automatically according to the UPPER value. For meter measurements, the range is fixed to the specified range.
Bandwidth switchingCan be expanded to a bandwidth that allows measurements from 0.1 Hz, which is required in the measurement of medical instruments and the like.
NormalNormal measurement bandwidth: 15 Hz to 1 MHz
ExpandExpands the measurement range to 0.1 Hz to 1 MHz
Total accuracy *2 (when network A, B, or C is used) *3Range1DC±(5.0 % of reading + 2 μA)
RMS0.1 Hz ≤ f < 15 Hz±(10.0 % of reading + 2 μA)
15 Hz ≤ f ≤ 100 kHz±(7.0 % of reading + 2 μA)
100 kHz < f ≤ 1 MHz±(10.0 % of reading + 2 μA)
Peak0.1 Hz ≤ f < 15 Hz±(10.0 % of reading + 10 μA)
15 Hz ≤ f ≤ 1 kHz±(10.0 % of reading + 10 μA)
1 kHz < f ≤ 100 kHz±(10.0 % of reading + 10 μA)
100 kHz < f ≤ 1 MHz±(20.0 % of reading + 10 μA)
Range2DC±(5.0 % of reading + 20 μA)
RMS0.1 Hz ≤ f < 15 Hz±(10.0 % of reading + 10 μA)
15 Hz ≤ f ≤ 100 kHz±(7.0 % of reading + 8 μA)
100 kHz < f ≤ 1 MHz±(10.0 % of reading + 10 μA)
Peak0.1 Hz ≤ f < 15 Hz±(10.0 % of reading + 10 μA)
15 Hz ≤ f ≤ 1 kHz±(10.0 % of reading + 10 μA)
1 kHz < f ≤ 100 kHz±(10.0 % of reading + 10 μA)
100 kHz < f ≤ 1 MHz±(20.0 % of reading + 10 μA)
Range3DC±(5.0 % of reading + 50 μA)
RMS0.1 Hz ≤ f < 15 Hz±(10.0 % of reading + 20 μA)
15 Hz ≤ f ≤ 100 kHz±(7.0 % of reading + 20 μA)
100 kHz < f ≤ 1 MHz±(10.0 % of reading + 20 μA)
Peak0.1 Hz ≤ f < 15 Hz±(10.0 % of reading + 50 μA)
15 Hz ≤ f ≤ 1 kHz±(7.0 % of reading + 50 μA)
1 kHz < f ≤ 100 kHz±(10.0 % of reading + 50 μA)
100 kHz < f ≤ 1 MHz±(20.0 % of reading + 50 μA)
Range4DC±(5.0 % of reading + 0.5 mA)
RMS0.1 Hz ≤ f < 15 Hz±(10.0 % of reading + 0.2 mA)
15 Hz ≤ f ≤ 100 kHz±(7.0 % of reading + 0.2 mA)
100 kHz < f ≤ 1 MHz±(10.0 % of reading + 0.2 mA)
Peak0.1 Hz ≤ f < 15 Hz±(10.0 % of reading + 0.5 mA)
15 Hz ≤ f ≤ 1 kHz±(7.0 % of reading + 0.5 mA)
1 kHz < f ≤ 100 kHz±(10.0 % of reading + 0.5 mA)
100 kHz < f ≤ 1 MHz±(20.0 % of reading + 0.5 mA)
Input resistance1 MΩ ± 1 %
Input capacitance200 pF or less (internal voltmeter input capacitance: 100 pF or less)
Common mode rejection ratio10 kHz or less: 60 dB or more, 10 kHz to 1 MHz: 40 dB or more
Offset cancel functionCancels up to 10 mA of the unnecessary current from measurements. OFF function available.

*1 Voltmeter band expansion is possible when network I is selected. *2 0.1 Hz ≤ f < 15 Hz is for when voltmeter band expansion (VoltMeter BandWidth) is set to Expand. Requires at least 120 second of test time. *3 A value converted to current for measurements using Network A, B, C or H with voltmeter accuracy of this product as the reference. If a network other than A, B, C or H is used, calculate as follows:
For Network D, E, or I, the ■ part of ±(□% of reading + ■A) is half the value.
For F, the ■ part is one-third the value.
For G, the ■ part is one-fourth the value.
For PCC-1 or PCC-2, the ■ part is 3.3 times the value.