When testing the insulation resistance testing of capacitive devices, applying the test voltage initially results in a low resistance measurement due to the charging current.
To eliminate the impact of this charging current when determining the lower limit, the wait time is set between applying the test voltage and starting the lower limit judgment.
The wait time can be set from 0.3 to 10 seconds (with 0.1 second resolution).