KPM1000_Current measurement function

ItemSpecifications
Measurement rangeResolution5 mA0.0001 mA
10 mA0.001 mA
20 mA
50 mA
100 mA0.01 mA
200 mA
500 mA
1A0.1 mA
2A
5A
10A1 mA
20 A
Allowable crest factor6
Accuracy*1 *2 *3Effective input rangeFull range1 % to 120 % of the range
45 Hz ≦ f ≦ 66 Hz, LPF: On or off±(0.1 % of reading + 0.1 % of range)
66 Hz < f ≦ 400 Hz, LPF: Off±(0.1 % of reading + 0.2 % of range)
400 Hz < f ≦ 1 kHz、LPF:OFF±(0.1 % of reading + 0.2 % of range)
1 kHz < f ≦ 5 kHz LPF:OFF±( 3 % of range)
Residual noiseWith the inputs shorted LPF: On or off0.5 % of range
One-year accuracy (accuracy up to 12 months after calibration)1.5 times the reading errors for the accuracy at 6 months
Auto rangeUP conditionsWhen the value exceeds 120 % of the range or when a peak over-range occurs
DOWN conditionWhen the value is less than 30 % of the range, the value is less than or equal to 90 % of the next lower range, and a peak overrange is not occurring

*1 Under the following conditions: within the effective input range, within six months after the last calibration, at a temperature of 23 °C ± 5 °C, sine wave, power factor of 1, and common mode voltage of 0 V. If the waveform is asymmetrical (waveforms such as half-wave rectified waveforms and full-wave rectified waveforms), errors will occur. * 2 When you are using the 5 mA range and the input is less than 10 % of the range, add (0.1 % of range). * 3 The peak current accuracy is defined for a sine wave whose frequency is between 45 Hz and 66 Hz. Accuracy: ± (0.1 % of reading + 3.1 % of range)