KPM1000_Power measurement function

ItemSpecifications
Measurement range (combination of the voltage and current ranges)Resolution750 mW0.01 mW
1.5 W0.1 mW
3 W
6 W
7.5 W
15 W1 mW
30 W
60 W
75 W
150 W0.01 W
300 W
600 W
750 W
1.5 kW0.1 W
3 kW
6 kW
Accuracy*1 *2Effective input rangeFull range1 % to 144 % of the range
45 Hz ≦ f ≦ 66 Hz, LPF: On or off±(0.1 % of reading + 0.1 % of range)
66 Hz < f ≦ 400 Hz, LPF: Off±(0.1 % of reading + 0.2 % of range)
400 Hz < f ≦ 1 kHz、LPF:OFF±(0.1 % of reading + 0.3 % of range)
1 kHz < f ≦ 3 kHz LPF:OFF±( 3 % of range)
Influence of the power factorPower factor 045 Hz to 66 Hz±0.4 % of VA
0 < Power factor < 1Add (tanθ × 0.4 )% of reading
One-year accuracy (accuracy up to 12 months after calibration)1.5 times the reading errors for the accuracy at 6 months

*1 Under the following conditions: within the effective input range, within six months after the last calibration, at a temperature of 23 °C ± 5 °C, sine wave, power factor of 1, and common mode voltage of 0 V. If the waveform is asymmetrical (waveforms such as half-wave rectified waveforms and full-wave rectified waveforms), errors will occur. *2 When you are using the 5 mA range and the input is less than 10 % of the range, add (0.1 % of range).


Related specifications : KPM1000