KPM1000_Voltage measurement function

ItemSpecifications
Measurement rangeResolution150 V0.01 V
300 V0.01 V
Allowable crest factor3
Accuracy*1 *2Effective input rangeFull range1 % to 120 % of the range
45 Hz ≦ f ≦ 66 Hz, LPF: On or off±(0.1 % of reading + 0.1 % of range)
66 Hz < f ≦ 400 Hz, LPF: Off±(0.1 % of reading + 0.2 % of range)
400 Hz < f ≦ 1 kHz、LPF:OFF±(0.1 % of reading + 0.2 % of range)
1 kHz < f ≦ 5 kHz LPF:OFF±( 3 % of range)
One-year accuracy (accuracy up to 12 months after calibration)1.5 times the reading errors for the accuracy at 6 months
Auto rangeUP conditionsWhen the value exceeds 120 % of the range or when a peak over-range occurs
DOWN conditionWhen the value is less than 30 % of the range, the value is less than or equal to 90 % of the next lower range, and a peak overrange is not occurring

*1 Under the following conditions: within the effective input range, within six months after the last calibration, at a temperature of 23 °C ± 5 °C, sine wave, power factor of 1, and common mode voltage of 0 V. If the waveform is asymmetrical (waveforms such as half-wave rectified waveforms and full-wave rectified waveforms), errors will occur. *2 The peak voltage accuracy is defined for a sine wave whose frequency is between 45 Hz and 66 Hz. Accuracy: ± (0.1 % of reading + 3.1 % of range)


Related specifications : KPM1000