Item/Model | 400W | 800W | 1200W | 2000W | |
---|---|---|---|---|---|
Weight (main unit only) | Approx. 3 kg (6.61 lb) | Approx. 5.5 kg (12.13 lb) | Approx. 7.5 kg (16.53 lb) | Approx. 13 kg (28.66 lb) | |
Dimensions | See the outline drawing.(Catalogue p15) | ||||
Environmental conditions | Operating environment | Indoor use, overvoltage category II | |||
Operating temperature | 0 °C to +50 °C (32 °F to +122 °F) | ||||
Operating humidity | 20 %rh to 85 %rh (no condensation) | ||||
Storage temperature | -25 °C to +60 °C (-13 °F to 140 °F) | ||||
Storage humidity | 90 %rh or less (no condensation) | ||||
Altitude | Up to 2000 m | ||||
Cooling method | Forced air cooling using fan | ||||
Grounding polarity | Negative grounding or positive grounding possible | ||||
Isolation voltage | L/ ML/ MH type: ±500 Vmax H type: ±800 Vmax | ||||
Withstanding voltage | Across the primary circuit and chassis | No abnormalities when 1500 Vac is applied for 1 minute | |||
Across the primary and secondary circuits | L/ ML/ MH type: No abnormalities when 1650 Vac is applied for 1 minute | ||||
H type: No abnormalities when 1900 Vac is applied for 1 minute | |||||
Across the secondary circuit and chassis | L/ ML/MH type: No abnormalities when 2300 Vdc is applied for 1 minute | ||||
H type: No abnormalities when 2640 Vdc is applied for 1 minute | |||||
Insulation resistance | Across the primary circuit and chassis | 100 MΩ or more (70 %rh or less) at 500 Vdc | |||
Across the primary and secondary circuits | LL/ ML/ MH type: 100 MΩ or more (70 %rh or less) at 500 Vdc | ||||
H type: 100 MΩ or more (70 %rh or less) at 1000 Vdc | |||||
Across the secondary circuit and chassis | L/ ML/ MH type: 40 MΩ or more (70 %rh or less) at 500 Vdc | ||||
H type: 40 MΩ or more (70 %rh or less) at 1000 Vdc | |||||
Accessories |
| ||||
Electromagnetic compatibility *1 *2 | Complies with the requirements of the following directive and standard.
Applicable under the following conditions: | ||||
Safety *1 | Complies with the requirements of the following directive and standard. Low Voltage Directive 2014/35/EU *2 EN 61010-1 (Class I *5 , Pollution Degree 2 *6) |
*1. Does not apply to specially ordered or modified products. *2. Limited to products that have a CE mark. Does not apply unless a core is attached to the J1 connector cable. *3. This is a Class A instrument. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *4. This is a Group 1 instrument. This product does not generate and/or use intentionally radiofrequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *5. This is a Class I instrument. Be sure to ground this product’s protective conductor terminal. The safety of this product is guaranteed only when the product is properly grounded. *6. Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only nonconductive pollution will occur except for an occasional temporary conductivity caused by condensation.