ItemCommon to all models
Synchronization function (clock synchronization)OverviewSYNC icon is displayed on the display when synchronization is established with the internal clock after connecting with other PXB series using the EXT SYNC connector.
Sequence synchronizationSynchronization of the program start and step start.
Measurement synchronizationSynchronization of the measurement start
Output synchronizationSynchronization of output ON/OFF
Sequence functionOperation modeCV, CC, and CP modes
Maximum number of programs30
Maximum number of steps10000
Step execution time1 ms to 3600 000 s
Loop count1 to 100000, or infinite
Sine FunctionOperation modeCV/CC mode
Frequency setting range1 Hz to 1000 Hz
Frequency resolution1 Hz to 10 Hz0.2 Hz
12 Hz to 100 Hz2 Hz
120 Hz to 1000 Hz20 Hz
CVMaximum settingSetting range up to 105 % of rated voltage
Maximum offset settingSetting range up to 105 % of rated voltage
CCMaximum settingSetting range up to 210 % of rated current
Maximum offset settingSetting range up to ±105 % of rated current
Pulse functionOperation modeCV/CC mode
Frequency setting range1 Hz to 1000 Hz
Frequency resolution1 Hz to 10 Hz0.01 Hz
12 Hz to 100 Hz0.1 Hz
120 Hz to 1000 Hz1 Hz
CVHigh level rated currentSetting range up to 105 % of rated voltage
Low level rated currentSetting range up to 105 % of rated voltage
CCHigh level rated currentSetting range up to 105 % of rated current
Low level rated currentSetting range up to 105 % of rated current
Duty cycle2.5 % to 97.5 %
Measurement triggerMeasurement start condition (trigger source)Conditions for starting measurement can be selected (when inputting from display, when inputting commands by remote control, when inputting signals by external control, and when operating in synchronization)
Number of measurements1 to 65536
Measurement delay timeSetting range0 s to 100 s
Resolution0.1 ms
Measurement intervalSetting range0.1 ms to 3600 s
Resolution0.1 ms
Measurement timeSetting range0.1 ms to 1 s
Resolution0.1 ms
I-V characteristic functionOperation modeCV/CC mode
Number of setup items3 to 100 items (interpolated between points with straight lines)
Preset memoryNumber of memory entries20
Saved settingValues in CV, CC, and CP modes, protection function values, and IR values
Setup MemoryNumber of memory entries21
Saved settingOn/off of the output from the DC OUTPUT terminal, Output voltage value/Output current value/Output power value, Output current for seamless operation (DC SEAM), Output mode, Response, Slew Rate, Priority operation mode (Priority when output is ON), Impedance Setting When the Output is Off (Impedance when output is OFF), Value of the pulse function (Duty, Frequency, High, Low), Value of the sine function (Amplitude, Frequency, Offset), Number of I-V characteristics (Count), Internal resistance value (IR), Over voltage protection (OVP), Under voltage protection (UVP, UVP Enable), Over current protection (OCP(+), OCP(-), Delay), Over power protection (OPP(+), OPP(-)), Line overvoltage protection (Line OVP), Measurement trigger settings (Source, Count, Delay, Enable, Timer), Integration settings (Gate, Reset)
key lockLevel 1Output on/off and preset memory recall are available
Level 2Output on/off are available
Level 3Output off is available
Number of parallel operated load modulesUp to 25 units