TOS5300_Other features / Interfaces

Item/ModelTOS5300TOS5301TOS5302
Double action featureTests can only be started by pressing and releasing STOP and then pressing START within 0.5 seconds of releasing the STOP switch.
Length of time to maintain a PASS judgment resultYou can set the length of time to maintain a PASS judgment: 50 ms, 100 ms, 200 ms, 1 s, 2 s,5 s, or HOLD.
Momentary featureTests are only executed while the START switch is held down.
Fail mode featureThis feature enables you to prevent remotely transmitted stop signals from clearing FAIL judgments and PROTECTION modes.
Timer featureThis feature finishes tests when the specified time elapses.
Output voltage monitor featureIf output voltage exceeds “setting + 350 V” or is lower than “setting – 350 V,” the TOS5200 switches to PROTECTION mode, output is turned off, and testing finishes.
MemoryUp to three sets of test conditions can be saved to memory.
Key lockLocks panel key operations (settings and changes).
Protective features Under any of the following conditions, the TOS5300 Series switches to the PROTECTION state, immediately turns output off, and stops testing. A message is displayed on the screen.
Interlock ProtectionAn interlock signal has been detected.
Power Supply ProtectionAn error was detected in the power supply.
Volt Error ProtectionWhile monitoring the output voltage, a voltage outside of the rated limits was detected. AC or DC withstanding voltage tests: ±350 V Insulation resistance test: ±(10 % of set + 10 V)
Over Load ProtectionDuring a withstanding voltage test, a value that is greater than or equal to the output limit power was specified. AC withstanding voltage test: 550 VA. DC withstanding voltage test: 55 VA.
Over Heat ProtectionThe internal temperature of the TOS5200 became too high.
Over Rating ProtectionDuring a withstanding voltage test, the output current was generated for a length of time that exceeds the regulated time.
Calibration ProtectionThe specified calibration period has elapsed.
Remote ProtectionA connection to or disconnection from the front- panel REMOTE connector was detected.
SIGNAL I/O ProtectionThe rear-panel SIGNAL I/O connector’s ENABLE signal has changed.
USB ProtectionThe USB connector has been disconnected while the TOS5300 Series was being controlled through the USB interface.
System clock Set in the following format: year/month/day hour/minutes/seconds.
Calibration dateSet when the TOS5300 Series is calibrated.
Calibration period settingSets the period before the next calibration is necessary.
Notification of when the calibration period elapsesSets the operation that is performed when the specified calibration period elapses. When the TOS5300 Series turns on, it can display a notification or switch to the protection mode and disable testing.
InterfaceUSBUSB Specification 2.0
REMOTEFront-panel 9-pin MINI DIN connector. By connecting an optional device to this connector, you can control the starting and stopping of tests remotely.
SIGNAL I/ORear-panel D-sub 25-pin connector