| Item | Specifications | ||
|---|---|---|---|
| Environment | Installation location | Indoors, 2000 m or less, Pollution Degree 2*1 | |
| Spec guaranteed range | Temperature | 5°C to 35°C (41°F to 95°F) | |
| Humidity | 20%rh to 80%rh (no condensation) | ||
| Operating range | Temperature | 0°C to 40°C (32°F to 104°F) | |
| Humidity | 20%rh to 80%rh (no condensation) | ||
| Storage range | Temperature | -20°C to 70°C (-4 °F to 158 °F) | |
| Humidity | 90 % rh or less (no condensation) | ||
| Power supply | Nominal voltage range (allowable voltage range) | 100Vac to 240Vac (90Vac to 250Vac) | |
| Power consumption | At no load (READY) | 100 VA or less | |
| Rated load | 800VA max. Maximum 200VA for TOS5403 only. | ||
| Allowable frequency range | 47Hz to 63Hz | ||
| Insulation resistance (between AC LINE and chassis) | 30 MΩ or more (500 Vdc) | ||
| Withstanding voltage (between AC LINE and chassis) | 1400Vac for 2 seconds | ||
| Earth continuity | 25 Aac/0.1Ω or less | ||
| External dimensions | 320(330)W × 132(150)H × 350(380)D mm(Maximum size) | ||
| Weight | TOS5401: Approx. 16kg (35.27lb), TOS5402: Approx. 15kg (33.07 lb), TOS5403: Approx. 9kg (19.84 lb) | ||
| Electromagnetic compatibility (EMC)*2 *3 | Complies with the requirements of the following directive and standards. EMC Directive 2014/30/EU EN 61326-1 (Class A*4) EN 61000-3-2 EN 61000-3-3 EN 55011 (Class A*4, Group 1*5) Applicable under the following conditions: The maximum length of all cabling and wiring connected to the TOS54 series must be less than 2.5m. Use the included high voltage test leads. Electrical discharges are applied only to the EUT. Shielded cables are being used when using the SIGNAL I/O. | ||
| Safety*2 | Complies with the requirements of the following directive and standards. Low Voltage Directive 2014/35/EU*3 EN 61010-1, EN61010-2-034 (Class I*6, Pollution Degree 2*1) | ||
*1 Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasional temporary conductivity caused by condensation.
*2 Does not apply to specially ordered or modified products.
*3 Only for models with CE marking / UKCA marking on their body.
*4 This is a Class A instrument. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
*5 This product belongs to Group 1 products. This product does not generate and/or use intentionally radio-frequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose.
*6 The TOS54 series is a Class I device. Be sure to ground the protective conductor terminal of the TOS54 series. The safety of this product is guaranteed only when the product is properly grounded.

