27813

TOS9300_General Specifications (Common)

ItemTOS9300TOS9301TOS9301PDTOS9302TOS9303TOS9303LC
Backup battery life3 years (at 25 °C or 77 °F)
EnvironmentInstallation locationIndoors, 2000 m or less
Spec guaranteed rangeTemperature5 °C to 35 °C (41 °F to 95 °F)(18 °C to 28 °C for partial discharge tests)
Humidity20 %rh to 80 %rh (20 %rh to 70 %rh for partial discharge tests)(no condensation)
Operating rangeTemperature0 °C to 40 °C (32 °F to 104 °F)
Humidity20 %rh to 80 %rh (no condensation)
Storage rangeTemperature-20 °C to 70 °C (-4 °F to 158 °F)
Humidity90 %rh or less (no condensation)
Power supplyNominal voltage range (allowable voltage range)100 Vac to 120 V, 200 V to 240 V (90 Vac to 132 V, 170 V to 250 V)
Power consumptionNo load(READY state)100 VA or less
Rated load800 VA max.
Allowable frequency range47 Hz to 63 Hz
Insulation resistance (between AC LINE and chassis)30 MΩ or more (500 Vdc)
Withstanding voltage (between AC LINE and chassis)1500 Vac, 1 minute, 20 mA or less
Earth continuity25 Aac, 0.1 Ω or less
WeightApprox. 17 kgApprox. 18 kgApprox. 22 kgApprox. 20 kgApprox. 21 kgApprox. 22 kg
AccessoriesPower cord (1 pc., *length: 2.5 m : The attached power cord varies depending on the shipment destination.), High-voltage test lead: TL31-TOS (1 pair), SIGNAL I/O plug (1 set), High-voltage warning sticker (1 pc.), Setup Guide (1 copy), CD-ROM (1 disc), Safety Information (1 copy), Heavy object warning label (1 pc., *Not included with the TOS9300), Test leads for earth continuity test: TL13-TOS (1 pair., *TOS9302, TOS9303,TOS9303LC only), [TOS9303LC only: Spare fuse (1 pc.), Test leads for leakage current test (2 red, 1 black), Flat probe (1 sheet)]
Electromagnetic compatibility *1 *2Complies with the requirements of the following directive and standards. EMC Directive 2014/30/EU EN 61326-1 (Class A *3), EN 55011 (Class A *3, Group 1 *4), EN 61000-3-2, EN 61000-3-3 Applicable under the following conditions The maximum length of all cabling and wiring connected to the product must be less than 2.5 m.Shielded cables are being used when using the SIGNAL I/O.The high-voltage test lead TL31-TOS is in use.Electrical discharges are applied only to the EUT.
Safety *1Complies with the requirements of the following directive and standards. Low Voltage Directive 2014/35/EU *2, EN 61010-1 (Class I *5 , Pollution Degree 2 *6)

*1 Does not apply to specially ordered or modified products. *2 Only on models that have CE/UKCA marking on the panel. *3 This is a Class A instrument. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *4 This is a Group 1 instrument. This product does not generate and/or use intentionally radio-frequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *5 This is a Class I instrument. Be sure to ground this product’s protective conductor terminal. The safety of this product is guaranteed only when the product is properly grounded. *6 Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non- conductive pollution will occur except for an occasional temporary conductivity caused by condensation.


Related specifications : TOS9300 Series

Please feel free to contact us with any questions or concerns.

Download a Product CatalogView List
Request a QuotationContact Us