Item | TOS9300 | TOS9301 | TOS9301PD | TOS9302 | TOS9303 | TOS9303LC | ||
---|---|---|---|---|---|---|---|---|
Backup battery life | 3 years (at 25 °C or 77 °F) | |||||||
Environment | Installation location | Indoors, 2000 m or less | ||||||
Spec guaranteed range | Temperature | 5 °C to 35 °C (41 °F to 95 °F)(18 °C to 28 °C for partial discharge tests) | ||||||
Humidity | 20 %rh to 80 %rh (20 %rh to 70 %rh for partial discharge tests)(no condensation) | |||||||
Operating range | Temperature | 0 °C to 40 °C (32 °F to 104 °F) | ||||||
Humidity | 20 %rh to 80 %rh (no condensation) | |||||||
Storage range | Temperature | -20 °C to 70 °C (-4 °F to 158 °F) | ||||||
Humidity | 90 %rh or less (no condensation) | |||||||
Power supply | Nominal voltage range (allowable voltage range) | 100 Vac to 120 V, 200 V to 240 V (90 Vac to 132 V, 170 V to 250 V) | ||||||
Power consumption | No load(READY state) | 100 VA or less | ||||||
Rated load | 800 VA max. | |||||||
Allowable frequency range | 47 Hz to 63 Hz | |||||||
Insulation resistance (between AC LINE and chassis) | 30 MΩ or more (500 Vdc) | |||||||
Withstanding voltage (between AC LINE and chassis) | 1500 Vac, 1 minute, 20 mA or less | |||||||
Earth continuity | 25 Aac, 0.1 Ω or less | |||||||
Weight | Approx. 17 kg | Approx. 18 kg | Approx. 22 kg | Approx. 20 kg | Approx. 21 kg | Approx. 22 kg | ||
Accessories | Power cord (1 pc., *length: 2.5 m : The attached power cord varies depending on the shipment destination.), High-voltage test lead: TL31-TOS (1 pair), SIGNAL I/O plug (1 set), High-voltage warning sticker (1 pc.), Setup Guide (1 copy), CD-ROM (1 disc), Safety Information (1 copy), Heavy object warning label (1 pc., *Not included with the TOS9300), Test leads for earth continuity test: TL13-TOS (1 pair., *TOS9302, TOS9303,TOS9303LC only), [TOS9303LC only: Spare fuse (1 pc.), Test leads for leakage current test (2 red, 1 black), Flat probe (1 sheet)] | |||||||
Electromagnetic compatibility *1 *2 | Complies with the requirements of the following directive and standards. EMC Directive 2014/30/EU EN 61326-1 (Class A *3), EN 55011 (Class A *3, Group 1 *4), EN 61000-3-2, EN 61000-3-3 Applicable under the following conditions The maximum length of all cabling and wiring connected to the product must be less than 2.5 m.Shielded cables are being used when using the SIGNAL I/O.The high-voltage test lead TL31-TOS is in use.Electrical discharges are applied only to the EUT. | |||||||
Safety *1 | Complies with the requirements of the following directive and standards. Low Voltage Directive 2014/35/EU *2, EN 61010-1 (Class I *5 , Pollution Degree 2 *6) |
*1 Does not apply to specially ordered or modified products. *2 Only on models that have CE/UKCA marking on the panel. *3 This is a Class A instrument. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *4 This is a Group 1 instrument. This product does not generate and/or use intentionally radio-frequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *5 This is a Class I instrument. Be sure to ground this product’s protective conductor terminal. The safety of this product is guaranteed only when the product is properly grounded. *6 Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non- conductive pollution will occur except for an occasional temporary conductivity caused by condensation.