TOS9300_Insulation Resistance Test-Other specifications

ItemTOS9300TOS9301TOS9301PDTOS9303TOS9303LC
Grounding mode (GND)Can be switched between Low and Guard.
LowGND is connected to the low terminal. Measures the current flowing across the low terminal and chassis (normal applications).
Guard *1GND is connected to Guard. Measures only the current flowing through the low terminal (current flowing through the chassis is not measured) (high sensitivity, high accuracy measurement applications).
Filter functionA low-pass filter can be inserted into the ammeter measurement circuit. *2

*1 If there is a possibility that the EUT or tools and the like will be grounded or if you are uncertain, do not set GND to Guard. Doing so is extremely dangerous because the ammeter will be shorted and will not be able to measure current. For normal applications, set GND to Low. *2 When the low pass filter is on, a judgment delay of at least 5 seconds and a test time are required.


Related specifications : TOS9300 Series