TOS9300_Other Functions (Common)

ItemTOS9300TOS9301TOS9301PDTOS9302TOS9303TOS9303LC
Auto testAuto execution by combining ACW, DCW, IR, and EC. For LC, a combination is possible only using TC, PCC, and Patient.
Test condition memorySetup MemoryUp to 51 test conditions (ACW, DCW, IR, EC, PD, LC) can be saved.
Program memoryUp to 100 program (ACW, DCW, IR, EC, PD) combinations, each containing 100 steps, can be saved.
Program memory (LC)Up to 100 program (TC, PCC, Patient) combinations, each containing 100 steps, can be saved.
Test result memoryRecords up to 1000 latest test result of independent tests and auto tests. These are cleared when the power is turned off. Test results can be saved in CSV format to a USB memory device.
System clock For recording the calibration time and test times
Recordable timeUp to year 2038
Calibration period settingDisplays a warning at power-on when the specified period passes. Select whether to activate a protection function or only display a warning in the display area when a warning occurs.
Display of measured values Maximum and minimum measurements can be displayed.
NormalDisplays measurements during a test. Maximum and minimum values are not held.
Maximum and minimum value displayDisplays the maximum current measurement for withstanding voltage (ACW/DCW) tests, the minimum resistance measurement for insulation resistance (IR) tests, the resistance measurement or voltage measurement for earth continuity (EC) tests.
Test start methodDouble ActionWhen you press STOP, “READY” is shown for 0.5 seconds. A test starts only when you press START within this period.
MomentaryTests are only executed while the START switch is held down.
Start LongA test starts only when the START switch is held down for at least 1 second.
PASS judgment display time (Pass Hold)Set the time to hold the pass judgment result display (0.05 s to 10.00 s) or hold it until STOP is pressed (Infinity).
STOP signal disable (Fail Mode)It is possible to set the instrument so that fail judgment results and PROTECTION mode cannot be released from a device connected to the SIGNAL I/O connector or REMOTE connector.
Key lockLock the operation of the keys to prevent changing the settings or overwriting memory or programs by mistake.
Protection functions If a protection function is activated during a test, the output is shut off and the test is stopped immediately.In an LC test, the power supply to the EUT is stopped, and the A and B terminals are opened.Conditions that cause a protection function to be activated are as follows.
INTERLOCKInterlock is activated.
POWER SUPPLYThere is an error in the power supply section.
OUTPUT ERRORAn output voltage outside of the following range is detected. ACW, DCW, IR test, PD test: ±(10 % of setting + 50 V) EC test: ±(10 % of setting + 2 A)
OVER LOADAn output power or output current outside of the following range is detected.
ACW: 550 VA, DCW: 110 W or 50 mA, IR (7200 V test): 110 W or 25 mA, IR (-1000 V test): 2 mA, EC: 240 VA, LC: AC LINE OUT current at approx. 15.7 A or power at 1600 VA.
OVER HEATThe internal temperature of the product is abnormally high.
OVER RATINGDuring a withstanding voltage test, an output current is generated for a length of time that exceeds the output time limit.
CALThe preset calibration period is exceeded.
REMOTEThe REMOTE connector is connected or disconnected.
SIGNAL I/OThere is a change in the SIGNAL I/O connector’s ENABLE signal.
COMMUNICATIONAn internal communication error is occurring.
OVER RANGEA value exceeding the maximum value of the measurement range is detected.
MEASUREAn error is detected in the LC test measurement check.
SHORTA relay operation error is detected in an LC test.
EARTH FAULTWhen the grounding mode (GND) is set to Guard, abnormal current flows from the high voltage output of this product to ground.
SCAN I/FWhile scanning, the interface cable is disconnected. Or, the channel-assigned scanner is not detected.

Related specifications : TOS9300 Series