AC output section | TOS9300 | TOS9301 | TOS9301PD | TOS9302 | TOS9303 | TOS9303LC | ||
---|---|---|---|---|---|---|---|---|
AC output section | Output range | 0.050 kV to 5.000 kV | ||||||
Resolution | 1 V | |||||||
Setting accuracy | ±(1.2 % of setting + 0.02 kV) (at no load) | |||||||
Max. rated load *1 | 500 VA(5 kV / 100 mA) | |||||||
Max. rated current | 100 mA (when the output voltage is 0.2 kV or higher) | |||||||
Transformer rating | 500 VA | |||||||
Output voltage waveform *2 | Sine | |||||||
Distortion | 2 % or less. (when the output voltage is 0.5 kV or higher and no load or a pure resistive load is connected) | |||||||
Crest factor | √2 ± 3 % (0.8 kV or higher) | |||||||
Frequency | 50 Hz / 60 Hz | |||||||
Accuracy | ±0.1 % | |||||||
Voltage regulation | ±3 % or less (when changing from maximum rated load to no load) | |||||||
Short-circuit current | 200 mA or more (output voltage 0.5 kV or higher) | |||||||
Output method | PWM switching | |||||||
Start voltage | The voltage at the start of the test can be set. | |||||||
Setting range | 1 % to 99 % of the test voltage | |||||||
Resolution | 1% | |||||||
Output voltage monitor function | If the output voltage exceeds ±(10 % of setting + 0.05 kV), the output is turned off, and the protection function is activated. |
27813
TOS9300_Withstanding Voltage Test-AC Output function
Related specifications : TOS9300 Series
- TOS9300_General Specifications (Common)
- TOS9300_Interface (Common)
- TOS9300_Other Functions (Common)
- TOS9300_Leakage Current Test-Other specifications
- TOS9300_Leakage Current Test-Timer function
- TOS9300_Leakage Current Test-Judgment function
- TOS9300_Leakage Current Test-Measurement section
- TOS9300_Leakage Current Test-Measurement circuit network
- TOS9300_Leakage Current Test-Measurement function
- TOS9300_Partial Discharge Test-Other specifications
- TOS9300_Partial Discharge Test-Timer function
- TOS9300_Partial Discharge Test-Judgment function
- TOS9300_Earth Continuity Test-Timer function
- TOS9300_Partial Discharge Test-Measurement function
- TOS9300_Partial Discharge Test-Output function
- TOS9300_Earth Continuity Test-Judgment function
- TOS9300_Earth Continuity Test-Measurement function
- TOS9300_Earth Continuity Test-Output function
- TOS9300_Insulation Resistance Test-Other specifications
- TOS9300_Insulation Resistance Test-Timer function
- TOS9300_Insulation Resistance Test-Judgment function
- TOS9300_Insulation Resistance Test-Measurement function
- TOS9300_Insulation Resistance Test-Output function
- TOS9300_Withstanding Voltage Test-Other specifications
- TOS9300_Withstanding Voltage Test-Timer function
- TOS9300_Withstanding Voltage Test-Judgment function
- TOS9300_Withstanding Voltage Test-Measurement function
- TOS9300_Withstanding Voltage Test-DC Output function
- TOS9300_Withstanding Voltage Test-AC Output function
- TOS9300_SIGNAL I/O Connector
- TOS9300_Programmable Detection Response Speed
- TOS9300_Test items