TOS9300_Withstanding Voltage Test-Other specifications

AC output sectionTOS9300TOS9301TOS9301PDTOS9302TOS9303TOS9303LC
Analog monitor *1 Outputs a voltage signal according to the current waveform or voltage waveform
ICurrent waveform: Scale 50 mA/1 V
VVoltage waveform: Scale 1 kV/1 V
Grounding mode (GND) Can be switched between Low and Guard.
LowGND is connected to the low terminal. Measures the current flowing across the low terminal and chassis (normal applications).
Guard *2GND is connected to Guard. Measures only the current flowing through the low terminal (cur-rent flowing through the chassis is not measured) (high sensitivity, high accuracy measure-ment applications).

*1 Monitor signal output is isolated from the chassis (earth). If you connect an oscilloscope or an external device whose BNC shield is grounded, be sure to set the grounding mode (GND) to Guard. The value is not calibrated. *2 If there is a possibility that the EUT or tools and the like will be grounded or if you are uncertain, do not set GND to Guard. Doing so is extremely dangerous because the ammeter will be shorted and will not be able to measure current. For normal applications, set GND to Low.


Related specifications : TOS9300 Series

Please feel free to contact us with any questions or concerns.