TOS9300_Withstanding Voltage Test-Timer function

AC output sectionTOS9300TOS9301TOS9301PDTOS9302TOS9303TOS9303LC
Current rise time settings range0.1 s to 200.0 s
Current fall time setting time *10.1 s to 200.0 s, OFF
Test time setting range0.1 s to 1000.0 s, OFF
Judgment delay (Judge Delay) setting range *20.1 s to 100.0 s, AUTO *3 (DCW only)
Accuracy±(100 ppm of setting + 20 ms) (excluding the fall time)

*1 This setting is used only when a PASS judgment occurs in ACW and DCW tests. During a DCW test, the voltage may not drop all the way within the set time because of the electrostatic capacity inside the product and the EUT. *2 Less than the sum of the rise time and fall time. *3 If Delay Auto is set to on, LOWER judgment is not made until the charge time ends.